Seiftest
FLUKE 900 SERVICE MANUAL
TEST 15
PULLPOS_FREQ
This test walks a positive pulse through the pull-ups latch to tiie frequency circuitry,
TESTS 16,17
TRIGQALO_EQUAL, TRIGQAL1_EQUAL
These tests walk a 0 or a 1 through the trigger and qualifier latches and the results are
determined through reading the EQUAL line.
TESTS 18,19
QAL0_EQUAL, QAL1_EQUAL
These tests walk a 0 or a 1 through the qualifier latches and the results are determined
through reading the EQUAL line.
NOTE:
For tests 18 and 19, an X in the general test result means that individual
results are imavailable.
TPQT^ Pn P1
TRIGO_EQUAL, TRIG1_EQUAL
These tests walk a 0 or a 1 through the trigger latches and the results are determined through
reading the EQUAL line.
NOTE:
For tests 20 and 21, an X in the general test result means that individual
results are unavailable.
TESTS 22 23
PULLO_TRIG_EQUAL, PULL1_TRIG_EQUAL
These tests walk a 0 or a 1 through the pull-ups latches, through the trigger circuitry and are
read through the EQUAL line.
NOTE:
For tests 22 and 23, an X in the general test result means that individual
results are unavailable.
TEST 24
PULLO TRIGQAL
This test walks a 0 through the pull-tq>s latches, through the trigger and qualifier circuitry and
the results are determined through reading the EQUAL line.
3-8