FLUKE 900 SERVICE MANUAL
Selftest
3.4 System Error Codes
System Error Codes appear on the command line of the 900 when its microprocessor operation
hangs up. These codes are used mainly for design debug, but are included here for completeness
in the event a software bug is encoimtered. The most common cause of such a system crash is the
loading of a corrupted simulated reference device file.
These codes are not of much use for hardware troubleshooting. The only hardware code is 80 and
its message spears similar to the following:
System Error #80
PC = 536A
Overlay = Selitest
Cannot clear Fault Latches
This indicates that at the time of the crash, the microprocessor Program Counter contained the
value 536A and memory addressing was in the Selftest overlay.
File Management Error Codes
01
FILE IS OPEN
02
DEVICE ALREADY FORMATTED
03
FILE NOT FOUND
04
FILE ALREADY EXISTS
05
INSUFHCIENT SPACE ON DESTINATION DEVICE
07
WRITTEN BYTE FAILED TO VERIFY
08
CANNOT WRITE DEVICE
09
FILE
CANNOT BE DELETED
OA DEVICE DOES NOT EXIST
OB FILE NOT OPEN
OC ERROR SETTING PAGE
OD ERROR SETTING CHAPTER
OE ERROR ACCESSING DEVICE
OF PASSWORD INVALID
10
INSUFnCIENT STACK SPACE FOR TRANSFER
11
UNABLE TO COMPRESS
12
CHECKSUM ERROR ON FILE
13
COPY PROTECTED
14
DEVICE NOT FORMATTED
15
FILE
UNRECOVERABLE/NOT FOUND
16
FILE IS DELETE PROTECTED
17
DEVICE HAS NO OVERHEAD
18
DIRECTORY EXCEEDS PAGE OVERHEAD
19
DCB DATA DOES NOT MATCH DEVICE DATA
lA ERROR IN OVERHEAD DATA
IB ADDRESS IS OUT OF FILE BOUNDS
1C FILE IS TOO BIG TO BE LOADED
ID DEVICE PASSWORD INCORRECT
IE END OF FILE REACHED
3 - 2 7