Service Manual Introduction
FLUKE 900 SERVICE MANUAL
©(restart) will re-execute the poweron selftest.
©(ignore) will bring up the main screen directly. If any of the failed selftests involve actual
testing circuitry (eg. FMxx), the operator will not be permitted to actually run a test cycle. File
manipulation, RD Test and System setting changes are possible, however.
Debug has three main keys: ©(run), ©(display), ©(prt_res). The last one means "print
results" and sends a formatted listing of all test results from the last test(s) executed to the serial
RS232 port, "display" brings to the screen, the characteristic 4 b)4e in^vidual result of the test
whose number was entered on the command line. This test must first have been run, either
explicitly or during the powerup.
The key labels under "run" are words that can be used to compose a command line instmction.
Some typical instmctions are:
Rim_test clip
Run test 0
These two equivalent instructions run the clip continuity test.
A clip must be inserted in the Input Bufter and clipped onto
the Dummy Chip Module which must be plugged into the
ZIF socket. Results are shown on the screen.
Run test 40
Test 40 is executed and result bytes displayed in the case of a
failure.
Run_test 1 thm 40
Run test continuous
Run_test 46 continuous
until_fault
Run_test 1 looping
Tests 1 through 40 are executed successively.
The series of poweron tests are executed repeatedly with the
number of passed and faUed cycles displayed on die screen.
will terminate it upon completion of the currently
executing test. This is a useful bumin procedure.
Test 46 is executed repeatedly while displaying the number of
passed cycles until the point at which a failure occurs.
Test 1 win execute repeatedly to facilitate signal tracing with
an oscilloscope.
1 - 2