Option -006

Limits

606-1. INTRODUCTION

606-2. The 21X0-006 Limits Option gives the 2180A
and 2190A Digital Thermometers the capability of
providing both visual and electrical indications (alarms)
when the temperature measurement exceeds either of a
pair of selectable maximum/minimum limit values. Also
selectable are, all for a single point, the delta function
(displays the difference between the preset value and the
temperature read), and the maximum/minimum display

function (the highest or lowest temperature read since the
last reset by the INITIALIZE MAX/MIN switch).

606-3. The Limits Option is available as factory
installed with initial order or may be ordered as a field
installable kit for addition to the instrument.

606-4. SPECIFICATIONS

606-5. Specifications for the Limits Option, 21X0-006,
are as listed in Table 606-1.

Table 606-1. Specifications

Limits Function: lights LED and activates Form A (SPST)

relay when a preset limit is exceeded. Contacts rated at

10 VA, 184V dc or 130V ac rms max, 0.5A max, resistive.

Min/Max Function: Store min or max readings, resettable

from front panel.

ATemperature Function: Reads ± temperature deviations

from preset nominal temperature.

Installable: Factory or field, through pre-punched front

panel.

Function:

Limit:

^ Low Limit

—9999°

>High Limit

+9999

Store min reading

Not used

Store max reading

Not used

Deviation from

Nominal

WARNING

H A Z A R D O U S V O L T A G E S M A Y B E
P R E S E N T W I T H I N T H E I N S T R U M E N T .
ONLY QUALIFIED PERSONNEL SHOULD
P E R F O R M T H I S I N S T A L L A T I O N P R O ­
CEDURE.

1.

Disconnect the thermometer from all power

sources.

2.

Remove the screws on the bottom of the case

that fasten the top and bottom of the PTI case
together and remove the top half of the case.

3.

Remove the hole plugs from the Limits section

of the front panel and attach the standoffs supplied
with the option to the front panel.

4.

Position the Limits PCB so that the

thumbwheel LED and pushbutton switch line up
with the applicable front panel ports.

5.

Attach the Limits PCB to the Main PCB using

the screws supplied with the option.

6.

Connect the cable on the Limits PCB to J4 on

the Main PCB.

7.

Replace the PTI cover on the instrument and

reconnect to input power sources, if required, at this
time.

606-8. OPERATING NOTES

606-6. INSTALLATION

606-7. Options for field installation can be installed
using the following procedure:

606-9. Installation of the Limits Option enables the low
current single contact relay (Kl) on the thermometer

Main PCB. The contact points are available on a rear

panel connector block.

606-1