beautypg.com

9. output connections – Fluke 2180A User Manual

Page 54

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

background image

21X0A-002

Table 602-2. Switch Selection

SWITCH

POSITION

SWITCH BANK

S1

BAUD RATE

S2

FUNCTION

S3

ADDRESS

0

1

2

3

4

5

6
7

8

9

OFF
ON

OFF
ON
OFF
ON

OFF
ON
OFF

ON

OFF
OFF

ON
ON
OFF
OFF

ON
ON
OFF
OFF

OFF
OFF

OFF
OFF
ON

ON
ON
ON
OFF

OFF

OFF
OFF

OFF
OFF
OFF
OFF

OFF
OFF

ON
ON

110
150

300
600

1200

2400

4800
9600

110
150

OPERATE
CAL 1
CAL 2
CAL3

CAL 4
CAL 1
CAL 2
CAL3
PLOT 1
PLOT 2

ADR 0
ADR 1
ADR 2

ADR 3
ADR 4

ADR 5
ADR 6
ADR 7

ADR 8
ADR 9

5.

Reinstall the Main PCB in the bottom half of

the case.

6. On the Output Unit PCB use Table 602-2 and S1
to select the desired BAUD rate, select position 0 on
the Function switch S2, and select the PTI Bus
Address using S3. When using the RS-232-C inter­
face, set the address switch S3 to address 1,6, 7, 8, or

9. When using the 2XXX A-522 Personality Card and
the 1120A IEEE-488 Translator, set the baud rate to
2400 and follow all procedures regarding the RS-232-
C interface.

7.

Connect the Output Unit cables to the

connectors on the Main PCB.

8.

Attach the Output Unit PCB to the spacers,

component side down, and the connector to the rear
panel access port.

9.

Replace the PTI cover on the instrument and

reconnect the input power sources, if required, at
this time.

602-9. OUTPUT CONNECTIONS

602-10. All connections between the Output Unit
Option and external instruments are made using the rear
panel connectors previously described. The analog
output is from standard banana jacks. The digital output
female connector has a mating male connector
accompanying the option. This allows the customer to
custom make a cable between the thermometer output

option and the receiving device. Table 602-3, is the pin­
out data for the digital output connector.

Table 602-3. Digital Output Connector Pin-Out

PIN

NO.

FUNCTION

MNEMONIC

USE

1

Address Valid

ADRVAL

PTI Bus

2

Data Valid

DATVAL

PTI Bus

3-6

Printer Address

A0-A3

PTI Bus

7

Acknowledge

ACK

PTI Bus

8

Not Used

9-16

Data

D0-D7

PTI Bus

17

Ground

PTI Bus

18

+5 Volts

PTI Bus

19-24

Not Used

30

Transmitted Data

RS232
1 nterface

31

Request to send

RS232
1 nterface

32

Clear to send

RS232
Interface

33

Data set ready

RS232
Interface

34

Signal Common

RS232
Interface

35

S0+

Current

Loop

36

SO-

Current

Loop

NOTE

Standard RS232C signals are output on 25-

pins, the connector on the -002 Output Option

is

36-pins,

therefore,

the

user

must

either

hardwire

the

connection

between

the

-002

Output Option or order the Y2026B, Cable

Adapter,

602-2