Page 70
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".
21X0A-004
Table 604-3. Message Format
CHARACTER POSITION
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
NORMAL TEMPERATURE MEASUREMENT DATA
CH 1
1
! CH
SP
SP
±
SP
D
D
D
D/.
D/.
D
SP
d
F/C
SP
!
SP
SP
CR
LF
OPEN THERMOCOUPLE OUTPUT (2190A only)
--------1
CH !
1
CH
SP
SP
±
SP
D
D
D
D/.
D/.
D
SP
d
F/C
SP
!
0
C
CR
LF
OVERLOAD OUTPUT
--------r
CH 1
____ i
CH
SP
SP
±
SP
D
D
D
D/.
D/.
D
SP
d
F/C
SP
!
0
L
CR
LF
SYMBOL
DESCRIPTION
CH
SP
±
D
D/.
d
F/C
!
OC
OL
CR
LF
Channel identification numbers (00 through 99,00 unless connected to Y2000, Y2001, or
2300A).
Space
Plus or minus symbol
Temperature data values (0 through 9) for the 2180A, position 12 for the 2190A.
Either a temperature data value or a decimal point
a lower case (d)
Fahrenheit or Celsius
Character (!) when the present limit of the Limits Option 21X0A-006 is exceeded.
One space if within limits.
Open (thermo)couple (character positions 18 and 19)
Overload (character positions 18 and 19)
Carriage Return
Line Feed
10 ON ERROR GOTO 250
20 ON CTRL/C GOTO 260
30 INIT PORT O
40 CLEAR PORT 0
50 TRIG @2
60 WAIT 2000
70 TIMEOUT 350
190 INPUT @2/A$
210 PRINT A*
220 GOTO 190
250 PRINT 'OOPS!!! ERROR #'iERR; ' ON LINE #'iERL
260 PRINT 'bye-bye'
270 END
!Wait for -004 to finish reset process
ISet IEEE timeout longer than reading del
!Request a reading
Figure 604-2. Non-SRQ Sample Program
604-4