21X0A-004
Table 604-1. Option 21X0A-004 Specifications
ENVIRONMENTAL
Operating Power
100, 200, 220, 240V ac ± 10% 50/60 Hz, 3VA. (The 12-volt
battery operation is not supported.)
Operating Temperature
40°C (Outside ambient of thermometer.)
OPERATIONAL CONTROL
Switch 3 (SW3)
OUT-OF-LIMIT-SIGNAL
Exclamation point transmitted with Option 21X0A-006
installed only; not with Y2002.
OUTPUT
Protocol and Connections
IEEE-488 standard
IEEE-488 Interface Function Capability Codes
IDENTIFICATION
FUNCTION
CAPABILITY
SHI
Source Handshake
Full
AHI
Acceptor Handshake
Full
T6
Talker
Serial Poll (not talk only)
L4
Listener
Not listen only (see note)
SRI
Service Request
Full
DTI
Device Trigger
Full (see 604-22. for function description)
604-14. Message Format
604-15. Table 604-3. contains the sequence of information
in the output data string. The sequence is the fixed length
format presented on the IEEE bus.
604-16. Sample Programs
604-17. The following sample programs show two differ
ent situations in which the controller receives temperature
readings from the thermometer. The program listed in Fig
ure 604-1 instructs the controller to asynchronously retrieve
a temperature reading after an SRQ has been sent from the
thermometer. The program listed in Figure 604-2 instructs
the controller to ask for a temperature reading at a specific
program step.
604-18. SRQ Program
604-19. The SRQ program, when typed into a 1720A,
1722A, or 1752A controller informs the controller what to
do when an SRQ is received from the thermometer. The
program allows the controller to perform other tasks until
the thermometer has a temperature reading. The thermom
eter sends a 66 hexadecimal value back to the controller in
response to the serial poll (SPL). If more than one ther
mometer or more than one other instrument (a maximum
of 14) is connected to the controller, the program distin
guishes which instrument is sending the SRQ and deter
mines priority if more than one SRQ is received at the same
time. After the controller receives the temperature reading
and prints it on the controller screen, the program ends the
SRQ handling procedure and waits for another SRQ.
NOTE
The following programs are written in Fluke
BASIC and operate on the 1720A, 1722A, and
1752A controllers. A modification of the SRQ
program may be required to conform to other
IEEE controllers.
604-20. The Non-SRQ Program
604-21. The sample program in Figure 604-2 inputs and
prints temperature readings without the use of an SRQ. A
2-second delay is required after the TRIG @2 statement
(line 50) to allow Option 21X0A-004 to finish the reset pro
cess before requesting readings from it. The TIMEOUT
delay (line 70) must be set greater than 333 ms, so the con
troller waits for the thermometer to provide the next read
ing to the INPUT statement.
604-22. Group Execute Trigger
604-23. The Option 21X0A-004 responds to the IEEE-
488 GET command by simply resetting the two micropi ^-
cessors (U5 and Ull) on the board. Resetting the microp-
604-2