Table 4-3. 2180A Performance Tests

2180A

OHM

SELECT

READING

90 DAY

1

YEAR

INPUT

SWITCH

“F

°c

°F

°c

°F

°C

39.650

(0)

-238.00

-150.00

.100

.054

.104

.057

100

ohm

212.030

385 Pt

572.0

300.0

.30

.17

.32

.19

345.210

1292.0

700.0

.44

.25

.49

.27

38.777

(1)

-238.00

-150.0

.044

.024

.049

.027

100

ohm

213.472

390 Pt

572.0

300.0

.25

.14

.28

.16

348.446

1292.0

700.0

.42

.24

.47

.26

38.679

(2)

-238.00

-150.00

.100

.055

.105

.058

100

ohm

213.929

3916 Pt

572.0

300.0

.28

.16

.30

.17

349.323

1292.0

700.0

.42

.24

.47

.26

38.612

(3)

-238.00

-150.00

.044

.024

.049

.027

100

ohm

214.135

392 Pt

572.0

300.0

.25

.14

.28

.16

349.909

1292.0

700.0

.39

.21

.43

.25

71.80

(4)

-67

-55.00

.26

.15

.27

.15

100

ohm

161.70

N1

212

100.0

.37

.21

.38

.22

219.00

347

175.0

.39

.22

.40

.23

6.201

(5)

-100

-73.3

.33

.19

.34

.19

10

ohm

14.778

CU

300

148.9

.35

.20

.37

.21

HI -

■0+ +0

■o+s

1 n

o C

2180A DIGITAL

DC VOLTAGE

VOLTAGE

I THERMOMETER

CALIBRATOR

DIVIDER

6-V

lO/xV RES

Y2022 or

EQUIVALENT

Figure 4-3. Calibration Connections

4-5