Table 4-3. 2180A Performance Tests
2180A
OHM
SELECT
READING
90 DAY
1
YEAR
INPUT
SWITCH
“F
°c
°F
°c
°F
°C
39.650
(0)
-238.00
-150.00
.100
.054
.104
.057
100
ohm
212.030
385 Pt
572.0
300.0
.30
.17
.32
.19
345.210
1292.0
700.0
.44
.25
.49
.27
38.777
(1)
-238.00
-150.0
.044
.024
.049
.027
100
ohm
213.472
390 Pt
572.0
300.0
.25
.14
.28
.16
348.446
1292.0
700.0
.42
.24
.47
.26
38.679
(2)
-238.00
-150.00
.100
.055
.105
.058
100
ohm
213.929
3916 Pt
572.0
300.0
.28
.16
.30
.17
349.323
1292.0
700.0
.42
.24
.47
.26
38.612
(3)
-238.00
-150.00
.044
.024
.049
.027
100
ohm
214.135
392 Pt
572.0
300.0
.25
.14
.28
.16
349.909
1292.0
700.0
.39
.21
.43
.25
71.80
(4)
-67
-55.00
.26
.15
.27
.15
100
ohm
161.70
N1
212
100.0
.37
.21
.38
.22
219.00
347
175.0
.39
.22
.40
.23
6.201
(5)
-100
-73.3
.33
.19
.34
.19
10
ohm
14.778
CU
300
148.9
.35
.20
.37
.21
HI -
■0+ +0
■o+s
1 n
o C
2180A DIGITAL
DC VOLTAGE
VOLTAGE
I THERMOMETER
CALIBRATOR
DIVIDER
6-V
lO/xV RES
Y2022 or
EQUIVALENT
Figure 4-3. Calibration Connections