A&D Weighing AD-4406 User Manual
Page 54

Page 52
AD-4406 Weighing Indicator
14.
14.
14.
14.
RS-422/RS-485, Relay Output (OP-03)
RS-422/RS-485, Relay Output (OP-03)
RS-422/RS-485, Relay Output (OP-03)
RS-422/RS-485, Relay Output (OP-03)
Replacing the RS-232C interface with this option, the RS-422/RS-485 interface can
connect up to 32 indicators and control them from a computer or a PLC.
The functions of the RS-422/RS-485 interface are common to RS-232C except the
signal system.
The relays output the result of comparison.
Solid-state-relay
Maximum voltage
50VDC
Maximum current
100 mA DC
Maximum resistance 8
W
Pin connections
Function
Pin No.
Signal name
Direction
Description
1
SDA
Output
Transmission A terminal
2
SDB
Output
Transmission B terminal
3
RDA
Input
Receive A terminal
4
RDB
Input
Receive B terminal
5
TRM
-
Terminator resistance (100
W)
RS-422
RS-485
6
SG
-
Signal ground
7
HI
Output
Relay output HI
8
OK
Output
Relay output OK
9
LO
Output
Relay output LO
Relay
output
10
COM
-
Relay output common
Adaptable connector
TM-BLA10 (an accessory)
An AC adapter (an accessory) must be connected to the option board when using the
RS-422 or RS-485 interface. If using relay output only, the AC adapter is not required.
Circuit
14. RS-422/485, Relay Output (OP-03)
*+ 1- .1 %1/
ᎏ᎐
᎖᎒
᎓᎖
Relay Output Terminals
RS-422 Output
RS-485 Output
RS-422 Input
RS-485 Input
5 5&$ 4 4&$ 64/
፼
፻Ꮂǡ
፻፵Ꮂǡ
፻Ꮂǡ
፻፵Ꮂǡ
፸፷፷ǡ
RS-422/RS-485 Terminals
5)
&%
DC 7V - 10V
AC adapter Jack
(AC adapter is an accessory of OP-03)
Note: Please confirm that the AC adapter type is correct
for your local voltage and receptacle type.
+
-