A&D Weighing AD-4406 User Manual
Page 43

AD-4406 Weighing Indicator
Page 41
Toward the Zero band
Preliminary and Free fall output are holding the off state.
Over/Under comparison starts. If F23 is set to 30, judgment starts when Free fall turns
on. The judgment is not latched and the output is according to the state at the time.
The Weighing completion relay is turned on if F24 is set to 2. The on time is set by F25.
The READY mark is blinking regardless of the settings of F24 and F25.
Start is may be accepted at this state.
Returns to Zero band
Over/Under and Weighing completion output is off.
The READY mark is turned on.
11.2.3. Setting the Parameters of Setpoint Comparison
Refer to "11.1.2. Setting the Upper/Lower Limit Values".
Zero band value is set at F26 of the F-Functions, and the value does not belong
to a specific code memory, but is used commonly.
Setting Order and Display for Setpoint Comparison and Simple batch
F22
MODE
Display
1
2
3
4
5
7 Setpoint
Setpoint
Final
Free fall
Over
Under
Comparison
Class
Fi
FF
Hi
Lo
1 Comparator
OK
HI
LO
Triangle
1
8 Setpoint
Setpoint
Final
Free fall
Preliminary
Over
Under
Comparison
Class
Fi
FF
Pr
Hi
Lo
2 Comparator
OK
HI
LO
Triangle
1
2
9 Setpoint
Setpoint
Final
Free fall
Full
Over
Under
Comparison
Class
Fi
FF
Fu
Hi
Lo
3 Comparator
OK
HI
LO
Triangle
1
2
10 Simple
Setpoint
Final
Free fall
Over
Under
Batch
Class
Fi
FF
Hi
Lo
1 Comparator
OK
HI
LO
Triangle
1
11 Simple
Setpoint
Final
Free fall
Preliminary
Over
Under
Batch
Class
Fi
FF
Pr
Hi
Lo
2 Comparator
OK
HI
LO
Triangle
1
2
12 Simple
Setpoint
Final
Free fall
Full
Over
Under
Batch
Class
Fi
FF
Fu
Hi
Lo
3 Comparator
OK
HI
LO
Triangle
1
2
11.2. Setpoint Comparison