A&D Weighing AD-4406 User Manual
Page 29

AD-4406 Weighing Indicator
Page 27
Current Loop Output
Item
Parameter
Description
* 0
Displayed data
1
Gross data
2
Net data
3
Tare data
F36
Output data
4
Gross data/Net data/Tare data
0
Stream mode
1
Manual mode
2
Auto print mode (+)
3
Auto print mode (+/-)
4
When accumulation, automatically output
F37
Output mode
* 5
No output
* 0
No delay
F38
Delay for continual data
1
2.0 seconds (F36=4, except F37=0)
0
600 bps
1
1200 bps
F39
Baud rate
* 2
2400 bps
bps: bit per second
*: Initial settings
Serial Interface
tem
Parameter
Description
* 0
Stream mode, command is not acceptable
1
Manual mode, command is effective
2
Auto print mode (+) , command is effective
3
Auto print mode (+/-), command is effective
4
When accumulation, automatically output
F40
Output mode
5
Command mode
0
No output
* 1
Manual, Fixed format
2
Auto, Fixed format
3
Manual, Format 1 (F34)
4
Auto, Format 1 (F34)
5
Manual, Format 2 (F35)
F41
Accumulated data output
at accumulated data
display
6
Auto, Format 2 (F35)
* 0
No delay
1
0.5 second
2
1.0 second
3
1.5 seconds
F42
Delay for continual data
4
2.0 seconds
* 0
Not used
F43
Command address
1
Used (Address selected at F06)
* 0
Approx. 1 second
F44
Time out
1
No limitation
* 0
CR, LF
F45
Terminator
1
CR
*:
Initial settings
6.2. F-Functions