A&D Weighing AD-4406 User Manual
Page 14

Page 12
AD-4406 Weighing Indicator
Dual Range
ω
ω
ω
For the range function, refer to “5.3. Weighing Range Function”.
5.2.2.
Specifying the Range and Unit
Step1
The range and unit of measure are displayed.
Range display :
SinGL : single range
dUAL : dual range
To change the range function, use the
key
Unit display
The active unit is displayed. The unit can be changed;
such as kg or lb. Calibration is performed using the unit
displayed.
Press the
key to select a unit.
Press the +/- key to select a unit for calibration. The unit for
calibration (first unit) is displayed and the alternate unit (second unit)
is blinking.
ENTER
The key to store the data displayed and proceed to the next step.
ESC
The key to proceed to the next step without changing the data.
5.2.3.
Specifying the Resolution, Decimal Point Position and Format
Step 2 The resolution will be displayed as d 0.1 with decimal point. The indicator displays
triangle 1 and the first unit selected at the previous step.
Specify the position of the decimal point with the
key. Specify the display
format (point or comma) with the +/- key. Specify the resolution with the
1
key.
The decimal point format for serial data output is selected using the F-function
settings.
Press the ENTER key to store the parameters and proceed to the next step.
When pressing the ESC key, regardless of what is displayed, the indicator will
proceed to the next step without changing the parameters.
5.2. Calibration Procedure