Exiting the calibration mode – A&D Weighing AD-4406 User Manual
Page 17

AD-4406 Weighing Indicator
Page 15
Weighing a Mass less than the Maximum Capacity
Step12
Set a mass value using the following keys.
key .... To select a digit.
key .... To select a value.
Step13
Place a mass equivalent to the displayed value on the
weighing unit. Proceed to step 15.
Weighing a Maximum Capacity Mass
Step14
Place a mass equivalent to the maximum capacity on
the weighing unit.
Step15
Press the ENTER key after the STABLE mark
turns on. Proceed to step 17.
ESC key.......... To proceed to step 17 without
changing the span parameter.
Caution Do not press the
ENTER
key while the STABLE
mark is off (detecting motion). Arrange the
condition using the F00 filter function.
Digital Input
Step16
Pressing the
+/-
key, a stored input voltage
parameter of the span is displayed in the unit
of mV/V. Adjust the input voltage using the
following keys.
key ............ To select a digit.
key ............ To select a value.
ENTER key ...... To store the span parameter and
proceed to step 17.
ESC key............ To proceed to step 17 without
changing the span parameter.
5.2.10. Exiting the Calibration Mode
Step17
Check the CALEnd display.
Use the following keys.
CAL key..... To store the parameters and display
CALoFF . Proceed to the next step.
ESC key .... To store the parameters temporarily.
Proceed to the CAL 0 display.
5.2. Calibration Procedure
Turn on STABLE mark
Span calibration
Maximum capacity
To step 14, step 16
To step 12
ZERO
TARE
New mass value
Mass of
maximum capacity
To step 14
To step 16
Place the displayed
mass on the
weighing unit
step 13
step 14
F 2
ENTER
Stored input voltage
New input voltage
mV/V
+/
+/
+/
+/-
--
-
ZERO
TARE
F 2
ENTER
mV/V
Turn display off