AD-4406 Weighing Indicator

Page 15

Weighing a Mass less than the Maximum Capacity

Step12

Set a mass value using the following keys.

key .... To select a digit.

key .... To select a value.

Step13

Place a mass equivalent to the displayed value on the

weighing unit. Proceed to step 15.

Weighing a Maximum Capacity Mass

Step14

Place a mass equivalent to the maximum capacity on

the weighing unit.

Step15

Press the ENTER key after the STABLE mark

turns on. Proceed to step 17.

ESC key.......... To proceed to step 17 without

changing the span parameter.

Caution Do not press the

ENTER

key while the STABLE

mark is off (detecting motion). Arrange the

condition using the F00 filter function.

Digital Input

Step16

Pressing the

+/-

key, a stored input voltage

parameter of the span is displayed in the unit

of mV/V. Adjust the input voltage using the

following keys.

key ............ To select a digit.

key ............ To select a value.

ENTER key ...... To store the span parameter and

proceed to step 17.

ESC key............ To proceed to step 17 without

changing the span parameter.

5.2.10. Exiting the Calibration Mode

Step17

Check the CALEnd display.

Use the following keys.

CAL key..... To store the parameters and display

CALoFF . Proceed to the next step.

ESC key .... To store the parameters temporarily.

Proceed to the CAL 0 display.

5.2. Calibration Procedure

Turn on STABLE mark

Span calibration

Maximum capacity

To step 14, step 16

To step 12

ZERO

TARE

New mass value

Mass of

maximum capacity

To step 14

To step 16

Place the displayed

mass on the

weighing unit

step 13
step 14

F 2

ENTER

Stored input voltage

New input voltage

mV/V

+/

+/

+/

+/-

--

-

ZERO

TARE

F 2

ENTER

mV/V

Turn display off