A&D Weighing AD-4406 User Manual
Page 53

AD-4406 Weighing Indicator
Page 51
data
Weight, result of comparison etc.
$CL
CLear previous settings. The UFC command parameters can not be
changed partially.
$WT
Displayed data
$GR
GRoss data
$NT
NeT data
$TR
TaRe data
$HD
Header of Gross/Net/Tare. Refer to CF06.
$UT
Weight UniT
$ST
STable/Unstable
$CP
Result of ComParison
$ID
ID number specified at F06
$DN
Data Number increments with each output automatically
$CD
CoDe memory number
$AN
Accumulation count
$TL
TotaL weight
$CM
CoMma
$CR
CR code (0Dh)
$LF
LF code (0Ah)
$DE
DElete the last parameter
$DL
Inserting DeLay time (0.1 second step) Example: DL10 : 1.0 second delay
strings
Output the specified strings, enclosed by single quotation (’).
’ itself is described using three single quotations; ’’’
Example: ’A & D’
Set data bits = 8 bits if using the 8-bit characters.
hexadecimal Control code of the printer etc, preceded by #.
2 characters preceded by # is hexadecimal code (0 - 9, A - F).
Example: #09
The #FF code can not be used because it is used for internal
control.
Example
SF1,$ID$DN$CR$LF$GR$UT$CR$LF
Serial output data format 1, ID number, data number, carriage return and line
feed, Gross weight, unit, carriage return and line feed.
SF2,’ Welcome to A & D’$LF’Total weight ’$AN$TL$LF$LF
Serial output data format 2, the strings ’ Welcome to A & D’ and line feed, the
strings ’Total weight ’ accumulation count, total weight, and 2 sets of line
feed.
13.4. UFC Command