5

OPERATING PROCEDURE FOR USING I MODE

Setting sensitivity

Select the sensitivity setting procedure according to the target condi-
tions.

To set sensitivity using a moving target (Fully-automatic

calibration)

1. Pass a target through the optical

axis while pressing the SET
button (3 seconds or more).

2. Confirm that the calibration

indicator (orange LED) flashes.

3. Release the SET button. The

calibration indicator (orange
LED) goes off.

To detect a minute color difference (2-point calibration)

1. With a target in place, press the

SET button and release it. The
calibration indicator (orange
LED) illuminates.

2. With the target removed, press

the SET button and release it.
The calibration indicator (orange
LED) goes off.

For target positioning (Positioning calibration)

1. With no target, press the SET

button and release it.
The calibration indicator (orange
LED) illuminates.

2. Place a target in the position

where it is to be stopped.

3. Press the SET button for 3

seconds or more and confirm
that the calibration indicator
(orange LED) flashes.

4. Release the SET button.

For stable detection unaffected by dust or dirt (Maximum

sensitivity setting)

1. Under the conditions shown in

the figure, press the SET button
for 3 seconds or more until the
calibration indicator (orange
LED) flashes.

2. Release the SET button.

1234

MODE

SET

N.O

N.C

OFF

DLY.

OUT TIM SET

Calibration
indicator

Press the button
for 3 seconds or
more.

1234

MODE

SET

N.O

N.C

OFF

DLY.

OUT TIM SET

Calibration
indicator

1. Press and release

the button with a
target in place.

2. Press and release

the button without
the target.

1234

MODE

SET

N.O

N.C

OFF

DLY.

OUT TIM SET

Calibration
indicator

Press the button
and release it.

With no target

When the sensitivity difference is insufficient:
“- - - - -” flashes for 3 seconds on the LCD display monitor after
calibration.

1234

MODE

SET

N.O

N.C

OFF

DLY.

OUT TIM SET

Calibration
indicator

Press the button
for 3 seconds or
more.

LCD display indication

The display is factory-set to show the received light intensity.

Received light
intensity (4 digits)

MODE

Press this
button once.

Setting value (4 digits)

* The setting value flashes
on and off.

Received light intensity display

Received light intensity is displayed in 4
digit numbers by defining the maximum
value as approximately 4000.

The maximum/minimum values vary depending on the fiber unit
characteristics.

Setting value display

The current setting value is displayed.
This display enables sensitivity
adjustment using the UP/DOWN button.

Changing the setting value

You can change the setting value in the “Setting value display”.

In setting value display mode

increases the sensitivity.

MODE

SET

MODE

SET

(The setting value is
decreased.)

decreases the sensitivity.

(The setting value is
increased.)

Reference: Holding down the UP/DOWN button quickly changes
displayed values.