Hardy Enviro Bench Scales User Manual
Page 24
![background image](/manuals/657661/24/background.png)
Section 6: Serial I/O
12/09
19
0596-0315-01
Rev
B
When display shows correct test weight press PROGRAM/ENTER
Displays “LOAd1”, “LOAd2”, “LOAd3”, or “LOAd4” then raw counts
Apply test weights to scale.
Fourth span point
Press PROGRAM/ENTER key
entered
Displays correct weight. Note: UNITS LED blinking
PROGRAM/ENTER
Press
2 key or internal switch S1
Displays microvolts per division (x.xxuPd)*
Press any front panel key to return to “CAL” prompt
Press OVER/UNDER/NEXT to display “donE”
Press PROGRAM/ENTER to display “SAvE”
Press PROGRAM/ENTER to display “SAvEd”
Scale returns to weigh mode
*The “
uPd”
(microvolts per division) must be one (1.0) or greater for proper operation.