Hardy Enviro Bench Scales User Manual
Page 22

Section 6: Serial I/O
12/09
17
0596-0315-01
Rev
B
Displays “LOAd1”, then raw counts
Apply test weights to scale. Press PROGRAM/ENTER key
Displays correct weight. Note: UNITS LED blinking
Press the 2 key or internal S1
Displays microvolts per division (x.xxuPd)*
Press any front panel key to return to “CAL” prompt
Press OVER/UNDER/NEXT to display “donE”
Press PROGRAM/ENTER to display “SAvE”
Press PROGRAM/ENTER to display “SAvEd”
Scale returns to weigh mode
*The “
uPd”
(microvolts per division) must be one (1.0) or greater for proper operation.