3 multipoint calibration, Multipoint calibration – Hardy Enviro Bench Scales User Manual
Page 23
Section 6: Serial I/O
12/09
18
0596-0315-01
Rev
B
5.3
Multipoint Calibration
Press and hold the PROGRAM/ENTER key until “SEtUP is displayed.
Or press internal switch S1 to display “SEtUP”
Press the OVER/UNDER/NEXT key until “CAL” is displayed
Press the PROGRAM/ENTER key to display security level “SL”
Press OVER/UNDER/NEXT to select proper security level
Press PROGRAM/ENTER
Raw counts displayed
No
New Zero?
Yes
Press the ZERO key
Scale zero’d. Raw counts become zero reference
ZERO
Press
PROGRAM/ENTER
Displays “CALAt2”, “CALAt3”,
Displays “CALAt1”, then
or “CALAt4”, then calibration weight
a calibration weight value
Enter test weight value. Most Significant digit blinks.
Increase digit with UNITS key. Decrease digit with 1 key.
Move flashing digit left using TARE key, or right using OVER/UNDER/NEXT key.
ZERO key clears display, PRINT moves decimal to right.
Next page