Working with the laboratory jaw crusher – FRITSCH PULVERISETTE 1, Model I premium line User Manual
Page 31
c
Hex wrench - size 6
d Feeler gauge strip
e Open-ended wrench - size 19
n Cross-head screwdriver
After mounting the crushing plates (17, 21), you need to check the gap
width between the crushing plates.
1.
Pull the source material container (8) out of the Jaw Crusher.
2.
Open the lid (7).
3.
Remove the 6 fastening screws of the belt guard (6) using a cross-
head screwdriver (n) and lift off the belt guard (6). The flywheel
(21) can be seen.
4.
Use the crank (4) to set the minimum grinding gap of 0.3mm.
SeeÄ Chapter 5.3.1 ‘Setting the gap width’ on page 35
5.
Turn the flywheel (22) to position 1.
Tools
Check
Working with the laboratory jaw crusher
- 31 -