beautypg.com

Working with the laboratory jaw crusher – FRITSCH PULVERISETTE 1, Model I premium line User Manual

Page 31

background image

c

Hex wrench - size 6

d Feeler gauge strip

e Open-ended wrench - size 19

n Cross-head screwdriver

After mounting the crushing plates (17, 21), you need to check the gap

width between the crushing plates.

1.

Pull the source material container (8) out of the Jaw Crusher.

2.

Open the lid (7).

3.

Remove the 6 fastening screws of the belt guard (6) using a cross-

head screwdriver (n) and lift off the belt guard (6). The flywheel

(21) can be seen.

4.

Use the crank (4) to set the minimum grinding gap of 0.3mm.

SeeÄ Chapter 5.3.1 ‘Setting the gap width’ on page 35

5.

Turn the flywheel (22) to position 1.

Tools

Check

Working with the laboratory jaw crusher

- 31 -

This manual is related to the following products: