Checkline CAP-TT01 User Manual
Page 23
Series TT01 Digital Cap Torque Testers
User’s Guide
22
The entered calibration point is too close to the previous point.
14 PASSWORDS
Two separate passwords may be set to control access to the Calibration section and to the menu and
other keys. To access the passwords setup screen, select Passwords from the menu. The display
appears as follows:
14.1 Calibration Password
Select Calibration from the sub-menu. The display appears as follows:
To set the password, select Enabled, then Set Password. Use the UP and DOWN keys to increment and
decrement the value, from 0 to 9999. When the desired value has been selected, press ENTER, then
ESC to exit the sub-menu.
14.2 Menu Key Password
If enabled, a password must be provided every time the MENU key is selected. Select Menu Key from
the sub-menu. Follow the same procedure as described in the previous sub-section.
14.3 Locking Out Other Keys
Other keys may be locked out individually. Select any combination of keys (MODE, ZERO, DATA) by
pressing ENTER in the Passwords sub-menu. Pressing a locked key will prompt the message “KEY
PROTECTED” and then revert to the previous screen.
14.4 Password Prompts
If passwords have been enabled, the following will be displayed when pressing the MENU key or
accessing the Calibration section:
CALIBRATION
COUNTER-CLOCKWISE
Load too close
to previous.
Please try again.
PASSWORDS
Calibration
Menu Key
Mode Key
Zero Key
Data Key
CALIBRATION PASSWORD
*
Disabled
Enabled
Set Password
(0000 – 9999)
5000