Series TT01 Digital Cap Torque Testers

User’s Guide


19

3. To escape the Calibration menu at any time, press ESCAPE. The display appears as follows:

Selecting “Cancel” will revert back to the Calibration setup. Selecting “Exit w/o saving” will return
to the menu without saving changes.

4. After the number of calibration points has been entered, press ENTER. The display appears as

follows:

5. Place the tester on a level surface free from vibration, then press ZERO. The tester will calculate

internal offsets, and the display appears as follows:

If failed:

6. The following screen appears after the offsets have been calculated:

Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Press ENTER.

CALIBRATION

NOT COMPLETE


Cancel

Exit w/o saving

CALIBRATION

CLOCKWISE

Attach necessary
weight fixtures,
then press ENTER.

CALIBRATION

OFFSET

Sensor failed

Analog failed

CALIBRATION

OFFSET

Sensor passed
Analog passed

CALIBRATION

OFFSET


Please wait…

CALIBRATION

OFFSET

Place torque sensor
horizontally, then
press ZERO.