beautypg.com

2 device decontamination, 8 light source exchange – BUCHI ELS Detector C-650 User Manual

Page 50

background image

C-650

Operation Manual, Version B

8 Troubleshooting

50

8.7.2

Device decontamination

Set the evaporation temperature to 100 °C and the gas pressure to 2 bar (29 psi).

Pump the appropriate solvent through the system at the rate of 1 mL/min. The solvent will be de-
termined by the nature of the samples that were previously analyzed by the detector. If you do not
know the nature of the sample, ethanol is a good choice. Do not use solvents that can potentially
corrode the device. Maintain the flow and temperature during 3 hours at least.

Clean the outside of the detector with a non-abrasive cloth. If necessary, a liquid such as
soapy water or ethanol can be used to remove stains or foreign material.

8.8

Light source exchange

In case the light source needs to be exchanged, please contact the BUCHI Customer Service.