beautypg.com

3 optional accessories, 2 technical data overview, Optional accessories – BUCHI ELS Detector C-650 User Manual

Page 12

background image

C-650

Operation Manual, Version B

3 Technical data

12

3.1.3

Optional accessories

Table 3.3: Optional accessories

Product

Order number

Flow splitter

11059007

3.2

Technical data overview

Table 3:4: Technical data

Dimensions (L×H×D)

250Ч450Ч550 mm

Connection voltage

115 VAC (±10 %)/60 Hz/1.8 A
230 VAC (±10 %)/50 Hz/1.7 A

Weight

18.5 kg

Overvoltage category

II

Pollution degree

2

Environmental conditions
Temperature
Altitude
Humidity

For indoor use only
5–40 °C
Up to 2000 m above sea level
Maximum relative humidity 80 % for temperatures up to
31 °C, and then linearly decreasing to 50 % at 40 °C

Detection

High sensitivity photomultiplier

Light source

Selected high efficiency blue LED

Temperature range

Ambient to 100 °C

Gas flow control

Manual and computer controlled (power down) nebulization
gas flow and patented auxiliary gas flow

Gas consumption

Less than 5 l/min

Eluent flow rate

0.1 mL/min to 5 mL/min

Device control

Microprocessor with stand alone manual keypad or PC
control

Operating parameters control

Liquid crystal digital panel

Signal drift

Less than 2 mV/30 min

Signal output

0–1 V (analog), RS-232 (digital)

Inputs

Remote Autozero (contact closure and TTL)
Remote Power down Mode (contact closure and TTL)

Power Down Mode

General
Standby
Cleaning

Zero control

Manual Autozero and remote Autozero