C-650

Operation Manual, Version B

5 Putting into operation

37

5.9.5

Column noise test

NOTE
When strongly retained compounds are slowly eluted from the column, excessive noise will be
observed.

To determine the column noise:

Turn off the pump and connect the column.

Restart the pump and allow the mobile phase to flow through the system. It is suggested that
you flush the column with a strong solvent for a few minutes before attaching it to the detector.

Set the gain to 12 and monitor the baseline for a few minutes. A suitable column will provide a
baseline that is 20–50 mV above the solvent baseline.

NOTE
If the mobile phase contains acidic modifiers (e.g. CF

3

COOH), disconnect the detector and wash

the Chromatography system for 12 h before starting to analyze unknown samples. This wash
should be performed after the column noise test is completed, but need not to be performed
after each analysis.