8 | Help with faults
BÜCHI Labortechnik AG
32/42
Operation Manual Interface I100
8.2
Error messages
Error message
Remedy
Pressure sensor is defective.
u
Contact BUCHI Customer Service.
Pressure sensor not calibrated.
u
Calibrate pressure sensor (see Chap
ter 6.3 "Calibrating the pressure sen
sor", page 26).
Data transmission error on RS485.
u
Check RS485 lead and make sure
that it is only connecting an interface
unit to a recirculating chiller.
Excess pressure in system.
u
Set pump to continuous mode until
the system pressure is reduced.
Error reading from or writing to memory
(EEPROM).
u
Contact BUCHI Customer Service.