4 technical data, 1 interface i-100, 2 power adapter – BUCHI Interface I-100 User Manual
Page 14: 3 ambient conditions, 4 materials

3 | Product description
BÜCHI Labortechnik AG
14/42
Operation Manual Interface I100
3.4
Technical data
3.4.1
Interface I100
Dimensions (W x H x D)
160 x 105 x 120 mm
Weight
700 g
Voltage
30 V DC
Power consumption
10 W
Solenoid valve power
supply
24 V
Measurement range
1400 – 0 mbar
Regulating range
1100 – 0 mbar
Measurement accuracy
± 2 mbar (after calibration at constant temperature)
Temperature compensa
tion
0.07 mbar/K
Hysteresis
Automatic or 1 – 500 mbar
Vacuum connection
GL14
Enclosure rating
IP21
Approval
CE
3.4.2
Power adapter
Input voltage
100 – 240 V
Output voltage
30 VDC
Power consumption
40 W
Frequency
50/60 Hz
Enclosure rating
IP20
Approval
CE/UL/CSA
3.4.3
Ambient conditions
Max. altitude above sea level
2000 m
Ambient temperature
10 – 40 °C
Maximum relative humidity
80 % for temperatures up to 31 °C
decreasing linearly to 50 % at 40 °C
The Interface I100 may only be used in indoor areas.
3.4.4
Materials
Component
Material
Pressure foil
Polyester
Housing
PBT
Venting valve
PEEK
Membrane of venting valve
FFKM
Venting loop connection
PPS