2 continuous mode – BUCHI Interface I-100 User Manual
Page 24

6 | Operation
BÜCHI Labortechnik AG
24/42
Operation Manual Interface I100
1
1
1
2
Fig. 14: Display during evacuation sequence
u
To increase the specified pressure while evacuation is in progress, press the P↑
button (2).
The system pressure is increased by a small amount and the pump switches to
Hold mode. The current pressure is maintained even if the specified setting has not
yet been reached.
u
To cancel Hold mode, press the H Off button (1).
The specified pressure previously set is reinstated.
u
To stop the distillation process, press the STOP button.
u
If necessary, depending on the venting setting (see Chapter 5.4.1 "Controller set
tings", page 20), press the STOP button again to vent the system.
6.2.2
Continuous mode
u
Press the Cont. button.
The vacuum pump is switched on and keeps running until the STOP or Man. but
ton is pressed. After an hour, the vacuum pump switches to Eco mode (see vac
uum pump operating instructions).