Help with faults | 8

Operation Manual Interface I­100

31/42

8

Help with faults

8.1

Faults, possible causes and remedies

Fault

Possible cause

Remedy

Device does not
work

Device is not connected to
power supply

u

Check power supply (see
Chapter 5.2 "Connections",
page 18
).

Valve or pump
frequently
switches over

System leaking

u

Check system for leaks (see
Rotavapor operating instruc­
tions).

u

If necessary, replace tubing
and/or seals.

Hysteresis setting too small

u

Select a larger hysteresis
(for terminal vacuums
greater than 700 mbar, set to
automatic hysteresis; see
Chapter 6.2.1 "Manual
mode", page 23
).

Valve does not
switch over

Valve coil dirty

u

Contact BUCHI Customer
Service.

Valve lead not plugged in

u

Plug in valve lead.

Valve not adjusted

u

Contact BUCHI Customer
Service.

Vacuum is not
achieved

Back­evaporation from the re­
ceiving flask

u

Empty receiving flask (see
Rotavapor operating instruc­
tions).

System leaking

u

Check system for leaks (see
Rotavapor operating instruc­
tions).

u

If necessary, replace tubing
and/or seals.

Water pressure of glass filter
pump too low

u

Increase water flow (see
pump operating instructions).

Vacuum pump too weak

u

Use suitably dimensioned
vacuum pump.