Help with faults | 8
Operation Manual Interface I100
31/42
8
Help with faults
8.1
Faults, possible causes and remedies
Fault
Possible cause
Remedy
Device does not
work
Device is not connected to
power supply
u
Check power supply (see
Chapter 5.2 "Connections",
page 18).
Valve or pump
frequently
switches over
System leaking
u
Check system for leaks (see
Rotavapor operating instruc
tions).
u
If necessary, replace tubing
and/or seals.
Hysteresis setting too small
u
Select a larger hysteresis
(for terminal vacuums
greater than 700 mbar, set to
automatic hysteresis; see
Chapter 6.2.1 "Manual
mode", page 23).
Valve does not
switch over
Valve coil dirty
u
Contact BUCHI Customer
Service.
Valve lead not plugged in
u
Plug in valve lead.
Valve not adjusted
u
Contact BUCHI Customer
Service.
Vacuum is not
achieved
Backevaporation from the re
ceiving flask
u
Empty receiving flask (see
Rotavapor operating instruc
tions).
System leaking
u
Check system for leaks (see
Rotavapor operating instruc
tions).
u
If necessary, replace tubing
and/or seals.
Water pressure of glass filter
pump too low
u
Increase water flow (see
pump operating instructions).
Vacuum pump too weak
u
Use suitably dimensioned
vacuum pump.