Radio configuration guide

80016835 Ver A

9

Propaq CS Vital Signs Monitor

This procedure is for Propaq CS monitors with software versions 3.7 or later.

To access the monitor service menu

1.

From the Main menu, press SETUP > MORE > MORE > SERVICE > YES > MORE >
MORE > RADIO on the main screen.

2.

Record the monitor IP address.

3.

Go to

Step 3

of

“Radio configuration tool user interface”

on page 11.

Set the NetName on a Propaq CS

Use this procedure to change the network name assigned to the monitor (assuming that
the current network name is one of the pre-set names listed in the monitor Change Net
Name menu). To set a custom network name, see

“Set an SSID using the radio

configuration tool”

on page 14.

To change the network name

1.

From the Main menu press SETUP > MORE > MORE > SERVICE > YES > MORE >
MORE > RADIO > CHANGE NET NAME to access the Change Net Name screen.

The current network name is highlighted.

Note

You cannot configure the Propaq CS radio though the monitor user interface. To
configure the radio on a Propaq CS monitor, follow the instructions in

“Radio

configuration tool user interface”

on page 11.

WARNING Do not attempt to change the network name unless you are a
qualified biomedical service engineer or a Welch Allyn employee. Changing the
monitor network name causes the monitor to re-start and seek to connect with
the FlexNet network corresponding to the new name.

WARNING Change the network name in a non-clinical environment only.

CHANGE NET NAME