1-5

Set-up And Operation

After a 15-second warm-up period, the light output will stabilize at its
normal intensity. Plug in the fiber optic bundle to stabilize light output.

1.

Plug the Stryker Light Source into a hospital-grade receptacle.

2.

Turn the light-intensity control knob to its minimum position. Plug the
fiber optic bundle into the illuminator port and connect the opposite
end to the instrument.

3.

Turn the power switch to the on position. The lamp will take 15
seconds to complete its warm-up cycle.

4.

Set light intensity to desired level.

5.

Turn unit off when not in use.

NOTE: Do not restart lamp immediately after turning off. The lamp
may fail to start. If this occurs, turn the power switch off and wait 15
seconds before restarting the Stryker Light Source.

CAUTION: Verify that the supply voltage matches the voltage
range indicated on the rear of the Stryker Light Source.

To prevent temporary blinding, turn power on only after the
fiber optic bundle is plugged in and the intensity control knob is
set to the minimum position.

The fiber optic bundle port is hot when fiber optic bundle is
removed.

On

Minimum Light Output

Maximum Light Output

Fiber Optic Bundle Port

Control Knob