WARNING: possible explosion hazard. Do not use in the
presence of flammable anesthetics. To ensure patient and
operator safety, read and understand this manual before using
the Stryker Light Source.
Do not remove cover. Electrical shock hazard due to high
internal voltage. There are no user serviceable parts inside
except for the lamp and fuse, which are accessible through the
bottom and back cover panels. Refer all service to Stryker.
Replace fuses as marked. See Fuse Replacement Section,
page 7.
The high intensity light at the front of the light source and at the
tip of the Fiber Optic Bundle may give rise to high
temperatures and bright light.
To minimize the risk of injury, avoid direct viewing or contact
with sensitive tissues.
Do not use for neonate transillumination. Erythema may result.
To prevent temporary blinding, always plug the fiber optic
bundle into the Stryker Light Source before turning the power
ON.
Total system risk current — including the patient applied
instrument — should not exceed 100 microamps. An isolation
transformer is required if total system risk current exceeds 100
microamps when accessories are interconnected.
Do not sterilize the Stryker Light Source. Refer to Cleaning
section for cleaning instructions.
Do not switch the Stryker Light Source ON and OFF in rapid
succession. Doing so will dramatically shorten lamp life. Allow
lamp to cool down 10 to 15 seconds between uses.