Troubleshooting Guide

SYMPTOM

PROBABLE CAUSE

SOLUTION

• Lamp is not illumi-

nating and fan is
operating.

• Lamp was not given

adequate time to “cool
down” before restarting.

• Turn unit off and wait 20 seconds

before turning back on.

• Light Source has over-

heated.

• Verify environmental tempera-

ture does not exceed specifica-
tions and allow to cool down for
10 minutes. Check to ensure
vents and fan exhaust are not
obstructed.

• Lamp has burned out.

• Replace lamp in accordance

with instructions in the USER
MAINTENANCE
section of this
manual. Use only GE Inspection
Technologies replacement lamp
part number SPL-600

SYMPTOM

PROBABLE CAUSE

SOLUTION

• Lamp is not illumi-

nating and fan is
NOT
operating...

• Power cord not securely

connected to light source
or AC source.

• Check for secure connections.

• AC source is not active.

• Use another power source or

restore power to the current
source.

• Power cord is damaged

(may cause intermittent
operation).

• Replace power cord.

• Light Source power

switch is not turned on.

• Turn switch to the ( I ) position.

• Lamp Drawer is not com-

pletely closed.

• Push Lamp Drawer closed and

lock in place by tightening finger
screw.

• One or both fuses have

blown.

• Replace both fuses in accordance

with instructions in the USER
MAINTENANCE
section of this
manual. If fuses continue to
blow, return unit to service center.

If the above suggestions do not solve the problem, call GE Inspection
Technologies Service Center listed in this manual.
DO NOT ATTEMPT
DISASSEMBLY! THERE ARE NO USER SERVICEABLE
COMPONENTS OTHER THAN LAMP.