Transportation, Adjustments, Specifications – Welch Allyn Vital Signs Monitor, Mounting Accessories - User Manual User Manual

Page 9: Transport/storage, Transportation adjustments specifications

background image

5

Do not leave dust, dirt, or finger prints on the lens
surface. They will prevent a clear observation of the
image.

Transportation

To transport the colposcope, fold the overhead arm into
its fully collapsed position as shown below.

Adjustments

The Tilt Brake of the colposcope can loosen with use. If
this occurs, tighten the set screw.

To tighten the Fine Focus Adjustment Knob, firmly grasp
each side of the knob and twist each side in the direction
indicated on the knob.

Specifications

Electrical:

120 V~ 60Hz, 2.5A

Weight:

100 pounds

Intermittent Operation:

Continuous Operation

Operating Temperature:

50°F to 95°F (10°C to 35°C)

Operating Humidity:

95%rh Max. (Non
Condensing)

Transport/Storage

Temperature:

-10°F to 140°F (-23°C to
60°C)

Transport/Storage

Humidity:

30 to 95%rh Non
Condensing

Atmospheric Pressure:

700hpa to 1060hpa

Classification:

Class I, IPXO (Ordinary
equipment with no special
protection against ingress of
water/fluids.

Fuse:

UL Listed 2.5A, 250Vac
Sloblo, Qty. 2

Welch Allyn Part No: 88428

Degree of Protection
against Electrical Shock:

Type B Light Source

CAUTION:

The Welch Allyn colposcope contains
sensitive electronic equipment. Handle
with care. Report any container damage
during transportation immediately upon
receipt.