Adjustments, Specifications, Transport/storage -10˚f to 140˚f (-23˚c to – Welch Allyn Vital Signs Monitor, Mounting Accessories - User Manual User Manual
Page 12: Transport/storage 30 to 95%rh non, Adjustments specifications

8
Adjustments
The Tilt Handle of the colposcope can loosen with use. If
this occurs, tighten the set screw.
To tighten the Fine Focus Adjustment Knob, firmly grasp
each side of the knob and twist each side in the direction
indicated on the knob.
To adjust the location of the Light Source on the post,
hold the Light Source with one hand and loosen the knob
on the post. Carefully slide the Light Source to the
desired location and re-tighten the knob firmly.
Specifications
Electrical:
120 V~, 60Hz, 2.5A
Weight:
40 pounds
Intermittent Operation:
Continuous Operation
Operating Temperature:
50°F to 95°F (10°C to 35°C)
Operating Humidity:
95%rh max. (Non
Condensing)
Transport/Storage
-10°F to 140°F (-23°C to
Temperature:
60°C)
Transport/Storage
30 to 95%rh Non
Humidity:
Condensing
Atmospheric Pressure:
700hpa to 1060hpa
Classification:
Class I, IPXO (Ordinary
equipment with no special
protection against ingress of
water.
Fuse:
UL Listed 2.5A, 250V
Sloblo, Qty. 2
Welch Allyn Part No. 88428
Degree of Protection
against Electrical Shock:
Type B Light Source
CAUTION:
The Welch Allyn colposcope contains
sensitive electronic equipment. Handle
with care. Report any container damage
during transportation immediately upon
receipt.
Tilt
Adjust
Set
Screw