ProBP 3400 and the target device lost connection

Possible cause

Corrective action and explanation

Unexpected loss in Bluetooth communication
The ProBP 3400 powered down

1.

Exit the target application.

2.

Power down the ProBP 3400

3.

Restart the ProBP 3400

4.

Relaunch the target application.

There is too much distance between ProBP 3400 and

the target device

Move the ProBP 3400 closer to the target device.

There is a problem with the target device

Refer to operating system or computer hardware

manuals for further troubleshooting.