Clean the device, Caviwipes or sani-cloth plus, 70 percent isopropyl alcohol – Welch Allyn Connex ProBP 3400 digital blood pressure device - User Manual User Manual
Page 54
![background image](/manuals/641619/54/background.png)
To remove the battery:
1. Remove the screw from the battery compartment door with a Phillips screwdriver.
2. Remove the battery compartment door.
3. Remove the battery from the battery compartment.
4. Remove the new battery from the shipping bag.
5. Using the polarized guidelines on the battery label, insert the battery, positive (+) side first, into
the battery compartment.
6. Replace the battery compartment door.
7. Secure the screw with the screwdriver.
If the ProBP 3400 will be unused for a long period of time, remove the battery prior to storing the
device.
Clean the device
WARNING Electric shock hazard. Before cleaning the device, disconnect the
power cord from the power source and the device.
WARNING Take care to prevent water or other fluid from entering any
connectors on the device. Should this occur, dry the connectors with warm air.
Check the accuracy of all operating functions.
Caution The device is not heat-resistant. Do not autoclave.
The following agents are compatible with the device:
•
CaviWipes
™
•
Sani-Cloth
®
Plus
•
70 percent isopropyl alcohol
•
10 percent chlorine bleach/90 percent water solution
Note
Clean the device on a routine basis, according to your facility's protocols and
standards or local regulations.
CaviWipes or Sani-Cloth Plus
Follow the cleaning agent manufacturer's guidelines.
1. Using CaviWipes or Sani-Cloth Plus, wipe the surface of the device to remove all debris.
2. Allow the device surface to dry for a minimum of 10 minutes before using the device.
70 percent isopropyl alcohol
Wipe the device with a clean cloth slightly dampened with 70 percent isopropyl alcohol.
48 Maintenance and service
Welch Allyn Connex® ProBP™ 3400 digital blood pressure device