Electrical specifications
Power requirements:
Input: 100-240VAC, 0.18A, 50-60Hz
Output: 5VDC, 0.5A
Degree of protection:
Type BF applied part
Safety classification:
Class II
Internally powered:
Battery, lithium-ion type, 3.7V, 2100mAh, 7.8Wh
Protection against the ingress of water:
IPXØ
Safety mode of operation:
Continuous operation
Standards:
This device complies with the following standards:
EN/IEC 60601-1
EN/IEC 60601-1-2
EN 1060-1:1996 Specification for non-invasive
sphygmomanometers - Part 1: General requirements
EN 1060-3:1997 Specification for non-invasive
sphygmomanometers - Part 3: Supplementary requirements for
electro-mechanical blood pressure measuring systems
FCC Subpart 15C
FCC ID #PI4411B, IC 1931B-BTM411
Environmental specifications
WARNING Fire and explosion hazard. Do not operate the device in the presence
of a flammable anesthetic mixture with air, oxygen, or nitrous oxide; in oxygen-
enriched environments; or in any other potentially explosive environment.
Caution Use the device within stated operating temperature ranges. The device
will not meet performance specifications if used outside these temperatures
ranges.
Operating temperature
10°C to 40° C (50° to 104° F)
Storage temperature
-20°C to 50° C (-4° to 122° F)
Operating altitude
-170 to 4877 m (-557 to 16,000 ft.)
Storage humidity
15 to 95% (noncondensing)