Guidance and manufacturer's declaration, Emissions and immunity information – Welch Allyn Connex ProBP 3400 digital blood pressure device - User Manual User Manual
Page 63
![background image](/manuals/641619/63/background.png)
Guidance and manufacturer's declaration
Emissions and immunity information
Electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or user of
the device should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all establishments other than
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following warning is
heeded:
WARNING This equipment/system is intended
for use by healthcare professionals only. This
equipment/ system may cause radio
interference or may disrupt the operation of
nearby equipment
a
. It may be necessary to take
mitigation measures, such as re-orienting or
relocating the device or shielding the location.
Harmonic emissions
IEC 61000-3-2
Class B
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
a
The device contains a 5-GHz orthogonal frequency-division multiplexing transmitter or a 2.4-GHz frequency
hopping spread-spectrum transmitter for the purpose of wireless communication. The radio is operated
according to the requirements of various agencies, including FCC 47 CFR 15.247 and R&TTE Directive (1995/5/
EC). The transmitter is excluded from the EMC requirements of 60601-1-2:2007, but should be considered
when addressing possible interference issues between this and other devices.
57