10 percent chlorine bleach/90 percent water solution

1. Wipe the device with a clean cloth slightly dampened with a 10 percent bleach and 90 percent

water solution.

2. Rinse with a clean cloth slightly dampened with clean water.
3. Allow the device surface to dry for a minimum of 10 minutes before using the device.

Clean the accessories

Wipe the NIBP hose and any reusable cuffs with a damp cloth moistened in a mild detergent

solution.

The same cleaning agents used to clean the device can be used on the stand and mounting

accessories.

Note

Clean the device on a routine basis, according to your facility's protocols and

standards or local regulations.

Directions for use

Maintenance and service 49