2 reporting commands, 3 the resetconfiguration command, 4 parameter setting commands – Rice Lake Farm Bars - IQ plus 390-DC User Manual
Page 29: Reporting commands, The resetconfiguration command, Parameter setting commands

EDP Commands
25
5.1.2
Reporting Commands
Reporting commands (see Table 5-2) send specific
information to the EDP port. These commands can be
used in both setup mode and normal mode.
5.1.3
The RESETCONFIGURATION Command
The RESETCONFIGURATION command can be
used to restore all configuration parameters to their
default values. Before issuing this command, the
indicator must be placed in setup mode or test mode
(press and hold setup switch for approximately four
seconds to enter test mode).
Sending this command, followed by the RS (RESET)
EDP command, is equivalent to using the DEFLT
function on the TEST menu. See Section 8.7 on
page 38 for more information about test mode and
using the TEST menu.
NOTE:
All load cell calibration
settings are lost when the RESETCONFIGURATION
command is run.
5.1.4
Parameter Setting Commands
Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 5-3 through Table 5-8).
Current configuration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:
command
Most parameter values can be changed in setup mode
only. Use the following command syntax when
changing parameter values:
command=value
where
value
is either a number or a parameter value.
Use no spaces before or after the equal (=) sign. If you
type an incorrect command, the display reads
??
.
Changes to the parameters do not take effect until you
exit setup mode.
For example, to set the motion band parameter to 5,
type the following:
MOTBAND=5D
Command
Function
DUMPALL
List all parameter values
VERSION
Write IQ plus 390-DC software version
RS
Reset the indicator
P
Write current displayed weight with units
identifier.
The P command returns the current
displayed weight value to the EDP port,
along with the units identifier. If the indicator
is in an underrange or overload condition,
the weight value is replaced with ######
(overload) or %%%%%% (underrange).
S
Write one frame of stream format
Table 5-2. EDP Reporting Commands
Command
Description
Values
GRADS
Graduations
1–999 999
ZTRKBND
Zero track band
OFF, 0.5D, 1D, 3D
ZRANGE
Zero range
1.9%, 100%
MOTBAND
Motion band
1D, 2D, 3D, 5D, 10D, 20D, OFF
OVRLOAD
Overload
FS+2%, FS+1D, FS+9D, FS
SMPRAT
Sample rate
15HZ, 30HZ, 3.75HZ, 7.5HZ
DIGFLTR1
DIGFLTR2
DIGFLTR3
Digital filtering
1, 2, 4, 8, 16, 32, 64
DFSENS
Digital filter cutout sensitivity
2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
DFTHRH
Digital filter cutout threshold
NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD
TAREFN
Tare function
BOTH, NOTARE, PBTARE, KEYED
Table 5-3. CONFIG EDP Commands