Internal count, Calibration procedure, Used to switch between internal count and a/d – Rice Lake Digi DI-28SS Bench Scale User Manual
Page 16: Raw data
6.2. Internal Count
This function is provided to check the internal count value from
the A/D (pre-amp)
To Enter Internal AD Count Mode
1. Press and hold down the [REZERO] key.
2. While holding , press the [
←
←
←
← DIGIT] key two times, the
[TARE] key once.
3. Then release the [REZERO] key.
In This Mode, Each Key Functions As Follows
[
↑↑↑↑ INCREASE] key
Used to switch between internal
count and A/D
Raw Data
[REZERO] key
Used to rezero the internal counts
[TARE] key
Used to exit from internal count mode
[14]
6.5.2. Calibration Procedure
PROCEDURE KEY
OPERATION
DISPLAY
Remarks
1. Enter calibration
mode.
[REZERO]
[
←
←
←
← DIGIT] [TARE]
[TARE]
888888
Press and hold
[REZERO] key
while pressing
[DIGIT], [TARE],
| 1
[TARE] then release
[REZERO] key.
2. Enter the minimum
display.
[
↑↑↑↑ INCR]
| 2
Press [
↑↑↑↑ INCR] key.
(Ex. 0.002)
[REZERO]
[ 00.000
∇
Press [REZERO] to
advance through
calibration
3. Enter the capacity
weight.
[
←
←
←
←] [ ↑↑↑↑ ] [ ↑↑↑↑ ]
[
↑↑↑↑ ] [ ↑↑↑↑ ] [ ↑↑↑↑ ]
[
↑↑↑↑ ]
[ 06.000
∇
Press [DIGIT] once
and [
↑↑↑↑ INCR] key six
times. (Ex. 6KG)
[REZERO]
〈 06.000
∇
Press [REZERO] to
advance through
calibration
4. Enter weight to be
used for calibration. (If
other than capacity.)
[
←
←
←
←] [ ↑↑↑↑ ] [ ↑↑↑↑ ]
[
↑↑↑↑ ] [ ↑↑↑↑ ] [ ↑↑↑↑ ]
[
↑↑↑↑ ]
〈 02.000
∇
Press [DIGIT] once
and [
↑↑↑↑ INCR] key six
times. (Ex. 2KG)
[REZERO]
CAL 0
Press [REZERO] to
advance through
calibration
5. Set zero point.
[REZERO]
CAL SP
6. Place weight on
platter from step 4.
[REZERO]
xxxxxx
When calibration is
complete scale shows
internal count.
7. Exit internal count
mode.
[TARE]
xx.xxx
Press [TARE] key to
return to weighing
mode.
8. Rezero scale place
weight on scale to test
calibration.
[REZERO]
[19]