beautypg.com

0 assembly and installation – ETS-Lindgren 94606-6 Current Probe User Manual

Page 37

background image

Assembly and Installation

|

37

5.0 Assembly and Installation

Before connecting any components, follow the

safety information in the ETS-Lindgren

Product Information Bulletin included with your

shipment.

Equipment Setup to Measure RF Current

Standing waves can exist on the test conductor under test at or near
the resonant frequency. Under these conditions, several
measurements taken along the line will provide a complete picture of
the RF current distribution and amplitude.

F

OR A

S

INGLE

C

ONDUCTOR

1.

Place the probe jaws around the conductor so that the conductor
passes through the center opening.

2.

Lock the jaws together.

F

OR A

T

WO

-C

ONDUCTOR

C

ABLE

To evaluate the common mode component of the noise current (the
net effect of the currents leaving and returning): Place the probe over
both conductors at the same time.

To measure the interference current in either conductor separately:
Place the probe over each wire individually.

F

OR

M

ULTI

-C

ONDUCTOR

C

ABLES

The probe will measure the net external effects of all the currents in the
conductors that pass through the center of the probe.