Specifications
91550-1
91550-2
91550-5
Maximum Power
Current
(400 Hz):
350 amps,
50 Hz–1500 Hz
225 amps
225 amps
Maximum Power
Voltage:
No limitation; subject to adequate conductor insulation
Internal Loading:
No Yes
No
Rated Ouptut Load
Impedance:
50 Ω
50 Ω 50
Ω
Sensitivity Under
Rated Load:
0.17 microampere
with 1 microvolt
sensitivity
receiver and
6 ohm transfer
impedance
1.0 microampere
with 1 microvolt
sensitivity
receiver and
1 ohm transfer
impedance, or
10 mV across
50 ohm load for
0.01 amp signal
1.0 microampere
with 1 microvolt
sensitivity
receiver and
1 ohm transfer
impedance, or
10 mV across
50 ohm load for
0.01 amp signal