Warning

Operators instructions
CombiPress™, type PFMx

The diaphragm

Don’t

No high

Use no

is very sensible

toutch

pressure

tools

and easy to destroy

cleaning

Factory guarantee is void on mechanical damage
of the diaphragm

Don’t touch the diaphragm

Always put on the protection cap if removing

the instrument from the application

Don’t use a high pressure cleaner on the dia-

phragm

Don’t use any tools on the diaphragm, only a

soft brush may be used.

Installation of 3A approved and EHEDG certified products:

1) Use only a 3A approved counter part.

2) The inspection hole should be visible and drained.

3) Mount the instrument in a self drained position.

4) Level the inner surface of the pipe with the counter part.

5) Welding's should be grinded to Ra= 0.8

Example
with 8126928

Example
with 8126916

Leakage indication hole

must be placed downwards

Fitted
incorrect

Cannot be
drained

WARNING

After installation and configuration

Check the leak tightness between the welding
sleeve and the instrument

Check the tightness of glands or M12 plugs.

Check the tightness of the instrument cover

Mounting

Tighten the connection with a torque of:

PFMN

G½A/G1A

20 Nm