Hardware Specifications
Dako Autostainer Plus
Dimensions
1.02m W x 0.69 m D x 0.61 m H 40” W x 27” D x 24” H
Weight
140 lb. (63.5 kg)
Electrical Requirements
120 V 110/120 V (±10%) 60 Hz (±2Hz)
220 V 220/240 V (±10%) 50 Hz (±2Hz)
Normal operating
temperature
18º C-26º C (64º F-79º F)
Slide rack capacity
1 – 12 glass slides
Total slide capacity
1 – 48 glass slides
Reagent capacity
64 different reagents (15 Ml/reagent vial)
Reagent dispense volumes
100, 150, 200, 400 and 600 µL
Reagent probe volume
capacity
100 µL minimum, 1200
Computer Controller
These are the minimum
features provided. Dako
reserves the right to change
the computer controller
specifications at any time.
NOTE: Computer specifications may vary.
Monitor 14”
color
Printer
LaserJet or Inkjet Printer
Surge Protector
120 VAC, 15 Amps, 50/60 Hz
Software Specifications
Operating System
Windows® 98, XP or equivalent.
Operating Logic
Designed to calculate the most time-efficient sequence to
complete a programmed staining run.
Protocol Logic
Flexible selection of steps including rinses. Maximum of 35
protocol steps (including rinse steps)
Idle Rinse
Buffer is applied to slides during a staining run to keep
specimens we when no reaction is occurring.
Rinse Default Settings
Buffer rinse every 30 minutes prior to the start of a run, when
delayed start is selected. Buffer rinse every 30 minutes during a
staining run. Water rinse every 60 minutes after completion of
run. Default setting for time and the amount of buffer dispensed
can be changed by a Dako representative upon request.
Incubation Time
Minimum time is 2 minutes.