beautypg.com

Cooper Instruments & Systems DFI 1650 Multi-Channel Digital Force Indicator User Manual

Page 42

background image

CF 126

39

Nov 2001

R6/W6

Read/write Units Label

Purpose

Reads or writes the channel's "DISPLAY. UNITS" menu item

Usage

"#aaccR6

↵" to read, "#aaccW6n↵" to write.

# is the 'pound' or 'hash' character (ASCII decimal 35).
aa is the two-character instrument address.
cc is the two-character channel number
n is the four-character units label.
↵ is the 'carriage return' character (ASCII decimal 13).

Example

"#0001W6CATS

↵"

Reply

When writing: "OK" or "ERROR".
When reading: a four-character string.
Changing this value doesn't change the mathematically scaling of the channel's values;
that can be changed by altering the "CALIBRATION DATA -> FULL-SCALE VALUE"
menu item and performing a re-calibration.


R7/W7

Read/write Full-scale Range

Purpose

Reads or writes the channel's "FULL-SCALE VOLT" or "FULL-SCALE CURR" menu
items.

Usage

"#aaccR7

↵" to read, "#aaccW7n↵" to write.

# is the 'pound' or 'hash' character (ASCII decimal 35).
aa is the two-character instrument address.
cc is the two-character channel number
n is the argument defined below.
↵ is the 'carriage return' character (ASCII decimal 13).

Argument

If the channel's Signal Type Jumpers are set for "voltage", the argument value can be 5
or 10 for 5V or 10V input.
If the channel's Signal Type Jumpers are set for "current", the argument value must be
20 for 4-20mA input.

Example

"#0001W710

↵"

Reply

When writing: "OK" or "ERROR".
When reading: an ASCll-floating-point value.
Changing this value will change the gain of the amplifier circuit. It is recommended that a
re-calibration of the transducer to the channel be performed to maintain calibration.


R8/W8

Read/Write Shunt Cal Value

Purpose

Reads or writes the channel's "SHUNT CAL VALUE" menu items.

Usage

"#aaccR8

↵" to read, "#aaccW8n↵" to write.

# is the 'pound' or 'hash' character (ASCII decimal 35).
aa is the two-character instrument address.
cc is the two-character channel number
n is the shunt calibration value in engineering units.
↵ is the 'carriage return' character (ASCII decimal 13).

Example

"#0001W8147.89

↵”

Reply

When writing: "OK" or "ERROR".
When reading: an ASCll-floating-point value.
Changing this value has no effect until the channel is re-calibrated to the transducer with
the Shunt Calibration method.


R9

Read Signal Type Jumper

Purpose

Read the setting of the Signal Type Jumper

Usage

"#aaccR9

↵"

# is the 'pound' or 'hash' character (ASCII decimal 35).
aa is the two-character instrument address.
cc is the two-character channel number
↵ is the 'carriage return' character (ASCII decimal 13).

Argument
Example

"# 0001R9

↵"

Reply

When reading: an ASCll-floating-point value.