Bio-Rad Laboratories Resources

Table 1 lists Bio-Rad resources and how to locate what you need.

Table 1. Bio-Rad resources.
Resource

How to Contact

Local Bio-Rad Laboratories

Find information and contacts on the Bio-Rad Laboratories

representatives

website by selecting your country on the Home page

(www.bio-rad.com). Find the nearest international office listed on

the back of this manual

Technical notes and literature

Go to the Bio-Rad Laboratories website (www.bio-rad.com).

Type a term in the Search box and select Documents to find links

to literature

Technical specialists

Bio-Rad Laboratories Technical Support scientists provide our

customers with practical and expert solutions. To find local

Technical Support on the phone, contact your nearest Bio-Rad

Laboratories office. For Technical Support in the United States

and Canada, call 1-800-424-6723 (toll-free phone), and select the

Technical Support option

Warranty

The TC20

automated cell counter and associated accessories are covered by a standard

Bio-Rad warranty. Contact your local Bio-Rad Laboratories office for the details of the warranty.

Safety Use Specifications and Compliance

A Warning! label in this manual warns you about sources of injury or harm, including risk of
electrical shock.

Warning! Do not attempt to repair or remove the outer case of the TC20 automated cell counter
or other accessories. If you open this instrument, you put yourself at risk for harm to body or equipment
from electrical shock.

This instrument has been tested and found to be in compliance with all applicable requirements
of the following safety and electromagnetic standards.

Table 2. Environmental conditions for safe operation of the TC20 automated cell counter.
Transient category

Installation categories II

Operating power

90–240 VAC, 1.2 A

Frequency

50/60 Hz

Electrical input

12 VDC, 2.5 A

Installation site

Indoor use only

Operating temperature

10–35°C

Maximum relative humidity

20–80%

Altitude

<2,400 m

Pollution degree

2

Degree of protection

IPX0

Note: Do not store or operate the unit near a sink as contact with water could cause electric shock.
Unit is heavy, do not store or operate it at the edge of a laboratory bench. During operation unit
must be in upright position.
Do not tip the unit over if the counting slide is inserted in the slide port; this could cause accidental
spillage of sample and contaminate the unit.
Use only Bio-Rad counting slides, using other manufacturers´ slides could damage the unit.

TC20 cell counter is factory calibrated, no further calibration is needed.