Experion automated electrophoresis system – Bio-Rad Experion DNA Analysis Kits User Manual
Page 27
Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com
23
Error Code
Message
Probable Cause
Recommended Action
50019
The run is already present in the
tree view
An attempt was made to import
a run that is already present
Check the tree view for the imported run
50021
Failed to save the run
Check permissions to write to the data folder
50022
Cannot change the
electrophoresis station
simulation mode because
an electrophoresis station is
currently running
The software is trying to toggle
between normal and simulation
mode, which cannot be done
while the electrophoresis station
is running
Stop the run to switch modes. To switch modes,
select Tools > Options > Advanced and then
select or deselect Instrument Simulation Mode
50024
Select a project before starting
a run
Select a project, then start the run
50033
Failed to connect to the database
See the Security Edition startup guide or
software Help file for requirements for connecting
to the database
50046
Aborting the run
Lid is opened during the run
Do not open the lid during the run; to stop the
run, click Stop Run
50047
Cannot close Experion because
electrophoresis stations are
acquiring data
Wait for the run to complete before exiting
50060
Failed to save protein names
Check permissions to write to the data folder
50063
Cannot start run -
Electrophoresis station lid open
Close the lid and start the run
50064
Cannot start run - no chip
detected
Confirm that the chip is primed and loaded
properly and that it is seated correctly in the
electrophoresis station. Then restart the run
50068
Failed to subscribe to
electrophoresis station control
events
Restart the computer and try again
Disable the firewall
50075
Cannot logoff because the user
is currently running experiments
Wait until the run is complete prior to logging off
50081
Cannot perform software OQ
because configuration files are
missing
Login as a Windows Administrator and reinstall
Experion software
Application needs to be run with
administrative privileges
Right-click the Experion software desktop
icon to launch software, and select Run As
Administrator
Attempt to start assay without
script
A script was not received by the
electrophoresis station
Restart Experion software
BsnReader has stopped working
Restart the computer and power off/on the
electrophoresis station
Cannot communicate with
channel power supplies
Contact Bio-Rad Technical Support
Experion Automated Electrophoresis System