4.3 Vortex Station
Error
Probable Cause
Recommended Action
No movement or sound
There is no power to the vortex station
Check the power cord connections and for AC power to the
outlet
Unplug the power cord and check both fuses
If all of the above are fine, contact Bio-Rad Technical Support
Vortex adapter is damaged
or does not move but motor
makes some noise, or if
solution splashes out of chip
wells after vortexing
Contact Bio-Rad Technical Support
4.4 Software Errors
Error Code
Message
Probable Cause
Recommended Action
50001
Login failed
The username and/or password
is invalid
Log in again using a valid username and
password
50004
Failed to create the data file
folder
Log in as a Windows Administrator and reinstall
Experion software
50005
Failed to create the packet file
folder
Log in as a Windows Administrator and reinstall
Experion software
50006
Cannot start a run because the
electrophoresis station is in use
Wait for a run to complete before starting a
new run
50007
Failed to locate the
electrophoresis station selected
Confirm the electrophoresis station and PC are
connected and try again
50008
Failed to open the assay selected
Restart the computer and try again
Reinstall Experion software
50009
Cannot find the script for the
assay selected
Reboot the computer and run the assay again; if
the problem persists, reinstall Experion software
50011
The selected assay cannot
be run on this electrophoresis
station
The software is in simulation
mode; an assay cannot be run
on a simulation electrophoresis
station
Deselect simulation mode to run the assay.
Select Tools > Options > Advanced and then
deselect Instrument Simulation Mode
50013
Failed to create the data file
Check permissions to write to the data folder
50015
Failed to locate the run selected
The run information appears in
the tree view, but the data file is
missing
Import the data file again by browsing to the data
file location
50016
No assay selected
Select an assay, then start the run
50017
Unexpected error occurred in the
application. Please check the log
file for details
Take the action indicated in the system log file