beautypg.com

1 electrophoresis station, Experion automated electrophoresis system – Bio-Rad Experion DNA Analysis Kits User Manual

Page 24

background image

4.1 Electrophoresis Station

Error

Probable Cause

Recommended Action

Green LED on the front
panel is not illuminated

There is no power to the electrophoresis
station

Check the connections to the PC and to the power outlet

Confirm that the power button is depressed, or in the “on”
position

Experion icon on the
software screen is grayed-
out

There is no communication between the
computer and the electrophoresis station

Check the connections to the PC, then reboot the PC.
Power on the electrophoresis station before launching
Experion software

IV check failure
(Experion software)

One or more wells are not filled correctly, and
the electrodes are not immersed

Fill all sample wells of the chip with blank or replicate samples

Ensure pipets are properly calibrated

Air bubbles are interfering with the electrical
contact in one or more of the wells

Stop the run, remove the chip, and use a clean pipet tip to
remove the sample and apply a new aliquot

When pipetting, insert the tip vertically and to the bottom of
the well. Dispense the liquid slowly. Do not expel air at the end
of the pipetting step. Dispense only to the first stop on the
pipet. Positive (reverse) pipetting may also be used

An electrode is loose or disconnected

Perform a conductivity test with the cleaning chip or
conductivity chip from the Experion validation kit (catalog
#700-7051). Add 850 µl water to the chip, place the chip in
the electrophoresis station, and close the lid. In Experion
software, select Tools > Diagnose Instrument and in the
Instrument Diagnostics dialog, select Conductivity

If the station passes the test, focus on the sources of error
above. If it fails, contact Technical Support with the results

The chip is not primed properly

Repeat the analysis using a new chip

Unreliable runs: the
electrophoresis station
aborts a run or generates
variable data

The computer power saver options are
causing the runs to abort (this is a problem
only with Experion software prior to version
3.0)

Upgrade to the newest version of Experion software

Confirm that all power saver and/or power options are turned
off before starting the assay. Do not let the computer “sleep”
during a run. If the computer goes to sleep, the run will abort

Sample preparation procedures are
inconsistent

Use the same reagents and follow the same procedures for
each sample and each run. Do not modify the protocol to
compensate for sample concentration or variability

Use calibrated pipets

Check that reagents and chips have not expired

The chip is not primed properly

Repeat the analysis using a new chip

An incompatible sample buffer was used

Ensure that the buffer used to solubilize samples falls within
the buffer compatibility guidelines for the assay (see the assay
instruction manual for details)

20

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

Experion Automated Electrophoresis System